46

To exit the Setup Mode without saving changes (355):

1. Press

"

to begin exiting Setup Mode.

DISPLAY READS X Enter ~ =CAL!

2. Press

to bypass Calibration Mode.

DISPLAY READS X Enter ~ =Stor

3. Press

to exit without saving changes.

DISPLAY READS X Enter ~ =Undo

4. Press

(

to undo changes.

DISPLAY READS X Enter ~ =End

5. Press

(

to complete exit.

DISPLAY READS X 0.00

Factory Default

Parameter 65001 and 65002 are available to return the Model 350 to factory settings. Parameter 65001 will reset
parameters including the calibration, while parameter 65002 resets all parameters accept the calibration.

To change the baud rate from the default value of 9600 to 4800 (350):

1. Key

in

or

DISPLAY READS X P65002~default-Cal

2. Press

1

once.

DISPLAY READS X Are U sure?~default=Enter

3. Press

1

once.

DISPLAY READS X Setup done~P65002~default-Cal

4. Press

1 1

To change the baud rate from the default value of 9600 to 4800 (355):

1. Key

in

or

!

DISPLAY READS X P65002~default-Cal

2. Press

(

once.

DISPLAY READS X Are U sure?~default=Enter

3. Press

1

once.

DISPLAY READS X Setup done~P65002~default-Cal

4. Press

"

( (