If the calibration weight value was entered before applying the test weight, Add ~ Load is displayed indicating that the

test weight should now be applied. Apply the test weight, press

1

or

(

and proceed to Exiting Calibration on

page 59.

If it is necessary to repeat the calibration process, press

"

at the Cal ~ Good? prompt and repeat the calibration

process.

Establishing span with a 100lb test weight:

1. Place a 100lb test weight on scale.

DISPLAY READS X Enter ~ Load ~ xx.xx

2. Enter

100.

DISPLAY READS X 100

3. Press

or

(

to establish span.

DISPLAY READS X 100.00

4. Pause for motion delay.

DISPLAY READS X Cal ~ Good? ~ 100.00


When making a significant change to the previous calibration, or when the calibration weight is less than 5% of full
scale capacity, ReCal ~ ??? will be displayed instead of Cal ~ Good? In this event it is recommended that the

calibration be performed a second time. Press

1

or

(

to recalibrate, or press

"

to skip recalibration.

Exiting Calibration

Once zero and span have been established, the newly acquired calibration information must be saved to non-volatile
memory before it will be realized in the Weigh Mode.

To exit and save calibration information:

1. Press

1

or

(

to accept calibration.

DISPLAY READS X Enter ~ =Stor

2. Press

1

or

(

to save calibration.

DISPLAY READS X Enter ~ =End

3. Press

1

or

(

to exit calibration.

DISPLAY READS X 100.00

To exit without saving calibration information:

1. Press

1

or

(

to accept calibration.

DISPLAY READS X Enter ~ =Stor

2. Press

.

DISPLAY READS X Enter ~ =Undo