Exiting calibration – Holtgreven GSE-355 I.S. User Manual
Page 108
99
If the calibration weight value was entered before applying the test weight, Add ~ Load is displayed indicating that the
test weight should now be applied. Apply the test weight, press
1
or
(
and proceed to Exiting Calibration on
page 59.
If it is necessary to repeat the calibration process, press
"
at the Cal ~ Good? prompt and repeat the calibration
process.
Establishing span with a 100lb test weight:
1. Place a 100lb test weight on scale.
DISPLAY READS X Enter ~ Load ~ xx.xx
2. Enter
100.
DISPLAY READS X 100
3. Press
or
(
to establish span.
DISPLAY READS X 100.00
4. Pause for motion delay.
DISPLAY READS X Cal ~ Good? ~ 100.00
When making a significant change to the previous calibration, or when the calibration weight is less than 5% of full
scale capacity, ReCal ~ ??? will be displayed instead of Cal ~ Good? In this event it is recommended that the
calibration be performed a second time. Press
1
or
(
to recalibrate, or press
"
to skip recalibration.
Exiting Calibration
Once zero and span have been established, the newly acquired calibration information must be saved to non-volatile
memory before it will be realized in the Weigh Mode.
To exit and save calibration information:
1. Press
1
or
(
to accept calibration.
DISPLAY READS X Enter ~ =Stor
2. Press
1
or
(
to save calibration.
DISPLAY READS X Enter ~ =End
3. Press
1
or
(
to exit calibration.
DISPLAY READS X 100.00
To exit without saving calibration information:
1. Press
1
or
(
to accept calibration.
DISPLAY READS X Enter ~ =Stor
2. Press
.
DISPLAY READS X Enter ~ =Undo