Establishing a span – Holtgreven GSE-355 I.S. User Manual
Page 107

98
4. Press
(
to establish zero.
DISPLAY READS X Enter ~ Load? ~ 0.00
5. Place a 100lb test weight on scale.
DISPLAY READS X Code03
6. Press
"
.
DISPLAY READS X First ~ Zero? ~ -0.26
7. Remove any load on the scale.
DISPLAY READS X First ~ Zero? ~ -0.42
8. Press
(
to establish zero.
DISPLAY READS X 0.00
9. Pause for motion delay.
DISPLAY READS X Enter ~ Load ~ 0.00
10. Place a 100lb test weight on scale.
DISPLAY READS X Enter ~ Load ~ xx.xx
11. Enter
.
DISPLAY READS X 100
12. Press
(
to establish span.
DISPLAY READS X 100.00
13. Pause for motion delay.
DISPLAY READS X Cal ~ Good? ~ 100.00
14. Press
(
to accept calibration.
Enter ~ =Stor
15. Press
(
to save calibration.
DISPLAY READS X Enter ~ =End
16. Press
(
to exit calibration.
DISPLAY READS X 100.00
17. Remove the calibration weight.
DISPLAY READS X 0.00
Establishing A Span
Once a zero reference has been established, the Model 350/355 displays Enter ~ Load and awaits the entry of a span
(test load) value. This value may be entered before or after the test load has been applied.