Holtgreven GSE-355 I.S. User Manual

Page 105

background image

96

DISPLAY READS X Fast ~ Cal~First ~ Zero? ~2640.

3. Press

!

! !

.

DISPLAY READS X Only ~ Zero? ~ 2640.

4. Remove any load on the scale.

DISPLAY READS X Only ~ Zero? ~ 2620.

5. Press

(

to establish zero.

DISPLAY READS X 00.

6. Pause for motion delay.

DISPLAY READS X Cal ~ Good? ~ 00.

7. Press

(

to accept calibration.

DISPLAY READS X Enter ~ =Stor

8. Press

(

to save calibration.

DISPLAY READS X Enter ~ =End

9. Press

(

to exit calibration.

DISPLAY READS X 00.

R

ESET

C

ALIBRATION

Cal Reset may be necessary when an over-load or under-load condition exists, preventing the completion of the
calibration process. Calibration Reset adjusts the zero and gain factors of the A/D amplifier to factory default values
for maximum sensitivity.

After performing a calibration reset, a complete recalibration is required. The effects of a calibration reset do not take
effect until the Model 350/355 recalibrated and calibration information has been saved.

If Code 02 (under-load) or Code 03 (over-load) is displayed during calibration, press

to perform a

calibration reset.

Reset Calibration Gain Factors Example (350):

1. Press

+

.

DISPLAY READS X Setup

2. Press

О

О

2

О

1

.

DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ xx.xx

3. Remove any load on the scale.

DISPLAY READS X First ~ Zero? ~ xx.xx

4. Press

1

to establish zero.

DISPLAY READS X Enter ~ Load? ~ 0.00