Holtgreven GSE-355 I.S. User Manual
Page 105

96
DISPLAY READS X Fast ~ Cal~First ~ Zero? ~2640.
3. Press
!
! !
.
DISPLAY READS X Only ~ Zero? ~ 2640.
4. Remove any load on the scale.
DISPLAY READS X Only ~ Zero? ~ 2620.
5. Press
(
to establish zero.
DISPLAY READS X 00.
6. Pause for motion delay.
DISPLAY READS X Cal ~ Good? ~ 00.
7. Press
(
to accept calibration.
DISPLAY READS X Enter ~ =Stor
8. Press
(
to save calibration.
DISPLAY READS X Enter ~ =End
9. Press
(
to exit calibration.
DISPLAY READS X 00.
R
ESET
C
ALIBRATION
Cal Reset may be necessary when an over-load or under-load condition exists, preventing the completion of the
calibration process. Calibration Reset adjusts the zero and gain factors of the A/D amplifier to factory default values
for maximum sensitivity.
After performing a calibration reset, a complete recalibration is required. The effects of a calibration reset do not take
effect until the Model 350/355 recalibrated and calibration information has been saved.
If Code 02 (under-load) or Code 03 (over-load) is displayed during calibration, press
to perform a
calibration reset.
Reset Calibration Gain Factors Example (350):
1. Press
+
.
DISPLAY READS X Setup
2. Press
О
О
2
О
1
.
DISPLAY READS X Fast ~ Cal~First ~ Zero? ~ xx.xx
3. Remove any load on the scale.
DISPLAY READS X First ~ Zero? ~ xx.xx
4. Press
1
to establish zero.
DISPLAY READS X Enter ~ Load? ~ 0.00