Note that if a setpoint activation method is set to
1
, disabling Auto Tare will also disable the activation
of that setpoint.
P169
Auto Tare Clear (Toggle)
Enabling this feature will cause the current tare value to be cleared to zero every time the indicator stabilizes
within
±5 graduations of gross zero
P171 Analog
(Toggle)
Enable or disable the optional analog output module. See Analog Output Setup on page 23 for all parameters
associated with the Analog Output Module.
P179 Count
(Toggle)
When enabled, the quantity mode becomes accessible via the
!
key. The quantity mode is identified by
the illumination of the QTY annunciator.
P200 Baud
(Selection)
Set the desired baud rate for the communication port.
P201 Data
Bits
(Toggle)
Select 7 or 8 data bits for the transmission.
P202 Parity
(Selection)
Select Odd, Even or None for the transmission parity.
P203
Stop Bits (Toggle)
Select 1 or 2 stop bits for communication port transmissions.
P204
Comm Handshake (Selection)
Select from None, Software (Xon/Xoff), Hardware (CTS/RTS), or Both.
P210 Send
(Selection)
Transmission Send options:
Choice Number
Selection
Name
Description
P210.00
Off
All transmissions disabled.
P210.01
Press
Sends transmission with [PRINT] key.
P210.02
Cont.
Sends transmissions continuously.
P210.03
Cycle
Send single transmission after weight is reached
and motion ceases. Must return display value
below 0.1% of F.S. to reset for next transmission.
P212 Send
Stability
(Toggle)
Enabling Send Stability will delay any transmissions until a no-motion condition exists.
P213
Transmit Selection (Selection)
Select desired print output (0 – 14). The transmission will be initiated by the selected print operation (P210) a
nd / or the Remote Key selection (P800). See page 53 for more details on the available preset transmits.
P250 RS-485 Multi-Drop Network (Toggle)