50

Note that if a setpoint activation method is set to

1

, disabling Auto Tare will also disable the activation

of that setpoint.

P169

Auto Tare Clear (Toggle)
Enabling this feature will cause the current tare value to be cleared to zero every time the indicator stabilizes
within

±5 graduations of gross zero

P171 Analog

(Toggle)

Enable or disable the optional analog output module. See Analog Output Setup on page 23 for all parameters
associated with the Analog Output Module.

P179 Count

(Toggle)

When enabled, the quantity mode becomes accessible via the

!

key. The quantity mode is identified by

the illumination of the QTY annunciator.

P200 Baud

(Selection)

Set the desired baud rate for the communication port.

P201 Data

Bits

(Toggle)

Select 7 or 8 data bits for the transmission.

P202 Parity

(Selection)

Select Odd, Even or None for the transmission parity.

P203

Stop Bits (Toggle)
Select 1 or 2 stop bits for communication port transmissions.

P204

Comm Handshake (Selection)
Select from None, Software (Xon/Xoff), Hardware (CTS/RTS), or Both.

P210 Send

(Selection)

Transmission Send options:

Choice Number

Selection

Name

Description

P210.00

Off

All transmissions disabled.

P210.01

Press

Sends transmission with [PRINT] key.

P210.02

Cont.

Sends transmissions continuously.

P210.03

Cycle

Send single transmission after weight is reached
and motion ceases. Must return display value
below 0.1% of F.S. to reset for next transmission.

P212 Send

Stability

(Toggle)

Enabling Send Stability will delay any transmissions until a no-motion condition exists.

P213

Transmit Selection (Selection)
Select desired print output (0 – 14). The transmission will be initiated by the selected print operation (P210) a

nd / or the Remote Key selection (P800). See page 53 for more details on the available preset transmits.

P250 RS-485 Multi-Drop Network (Toggle)