Etup, Rrors, Ardware – Holtgreven GSE-355 I.S. User Manual
Page 116
![background image](/manuals/657678/116/background.png)
107
S
ETUP
M
ODE
E
RRORS
Bad ~ Code!
An incorrect access code was entered.
Unit ~ Seald
Access to the Setup or Calibration Mode was denied. Check the internal
“YES/NO” program jumper.
Unit3-Ntep
Parameter 440 (NTEP) is enabled and parameter 152 (third unit) is set to an
Code 49.
additional unit. The third unit is not NTEP approved.
Entry ~ Error
An invalid entry was made.
Need ~ Entry
A numeric value was required before pressing
#
.
Out of ~Range
The entered value exceeded the allowable range.
Can’t ~ Set!
Attempt to change a parameter that does not allow an entry.
ResGT ~ 260E3
The number of divisions exceeds 260000 (see P110, P111).
Code 35.
ResGT ~ 25E3
The number of divisions exceeds 25000 (see P110, P111).
ResLT ~ 100!
The number of divisions is less than 100 (see P110, P111).
ResLT ~ 1 !!
Number of divisions is less than one (see P110, P111).
SPtxx ~ Error
A conflict occurred with a setpoint value entry (example: target entry is less than
preact). The digits ‘xx’ represent the last two digits of the setpoint parameter
in error (example: SPt 5 ~ Error indicates a conflict P5105, preact 1).
Prtcl ~ Error
Existing protocol is invalid. The following are not allowed:
- P201 = 7 data bits, P202 = no parity, P203 = 1 stop bit
- P201 = 8 data bits, P202 = even parity, P203 = 2 stop bits
- P201 = 8 data bits, P202 = odd parity, P 203 = 2 stop bits
H
ARDWARE
E
RRORS
Code00
An EPROM problem detected during power-up (U2).
A-D ~ Bad!
Or Code17
Problem with A/D chip detected. Disconnect any options installed and re-
power the unit. Options are connected to the same serial lines as the A/D so
they may prevent it from working properly.
Deflt ~ A-D
Bad A/D calibration values. Recalibrate A/D (see A/D Calibration Procedure
on page 111.
Re- ~ Boot!
EEPROM data could not be read. Attempting power-up reset.
Chec ~ E2
EEPROM data error (U4).
Deflt ~ Setup
An error occurred when reading setup data from the EEPROM during power-
up. All parameters are set to factory default.
Ch.XXXX
A checksum error occurred during power-up. All anunciators are lit. The
EPROM integrity test failed or is improperly seated.
E2 ~ Full!
The EEPROM setup exceeds the memory capacity.
NoSpc ~ Free!
The current setup exceeds the setup RAM capacity.