beautypg.com

Etup, Rrors, Ardware – Holtgreven GSE-355 I.S. User Manual

Page 116

background image

107

S

ETUP

M

ODE

E

RRORS

Bad ~ Code!

An incorrect access code was entered.


Unit ~ Seald

Access to the Setup or Calibration Mode was denied. Check the internal
“YES/NO” program jumper.


Unit3-Ntep

Parameter 440 (NTEP) is enabled and parameter 152 (third unit) is set to an

Code 49.

additional unit. The third unit is not NTEP approved.


Entry ~ Error

An invalid entry was made.

Need ~ Entry

A numeric value was required before pressing

#

.


Out of ~Range

The entered value exceeded the allowable range.


Can’t ~ Set!

Attempt to change a parameter that does not allow an entry.


ResGT ~ 260E3

The number of divisions exceeds 260000 (see P110, P111).

Code 35.

ResGT ~ 25E3

The number of divisions exceeds 25000 (see P110, P111).


ResLT ~ 100!

The number of divisions is less than 100 (see P110, P111).


ResLT ~ 1 !!

Number of divisions is less than one (see P110, P111).

SPtxx ~ Error

A conflict occurred with a setpoint value entry (example: target entry is less than
preact). The digits ‘xx’ represent the last two digits of the setpoint parameter
in error
(example: SPt 5 ~ Error indicates a conflict P5105, preact 1).

Prtcl ~ Error

Existing protocol is invalid. The following are not allowed:
- P201 = 7 data bits, P202 = no parity, P203 = 1 stop bit
- P201 = 8 data bits, P202 = even parity, P203 = 2 stop bits
- P201 = 8 data bits, P202 = odd parity, P 203 = 2 stop bits

H

ARDWARE

E

RRORS

Code00

An EPROM problem detected during power-up (U2).

A-D ~ Bad!
Or Code17

Problem with A/D chip detected. Disconnect any options installed and re-
power the unit. Options are connected to the same serial lines as the A/D so
they may prevent it from working properly.

Deflt ~ A-D

Bad A/D calibration values. Recalibrate A/D (see A/D Calibration Procedure
on page 111.

Re- ~ Boot!

EEPROM data could not be read. Attempting power-up reset.

Chec ~ E2

EEPROM data error (U4).

Deflt ~ Setup

An error occurred when reading setup data from the EEPROM during power-
up. All parameters are set to factory default.

Ch.XXXX

A checksum error occurred during power-up. All anunciators are lit. The
EPROM integrity test failed or is improperly seated.

E2 ~ Full!

The EEPROM setup exceeds the memory capacity.

NoSpc ~ Free!

The current setup exceeds the setup RAM capacity.