BASIC OPERATION

75

4. Select the trace math function:

PowerDiff[F3]
LogDiff[F4]
LogOffset[F5]

5. If LogDiff was selected, set the reference level

and unit.

LogDiff ref range:
LogDiff ref units:

-120dBm ~ 30dBm
dBm, W

6. If LogOffset was selected, set the offset level

and unit.

LogOffset range:

-50dB~+50dB

7. To turn trace math off, press the OFF[F6].

Display Icon

The Math icon is displayed when trace
math is turned on.

Trace Detection Mode

Description

Each time the spectrum analyzer samples data
for each point on the trace, a number of samples
are usually taken for each point, known as a
sample bucket. The actual value of each point is
determined by the detector from the samples in
each bucket.

Each selected trace, (1, 2, 3, 4), can use a
different detection mode.