BASIC OPERATION

67

Operation

1. Press

Sweep

> Sweep Time[F1] and toggle the

Sweep time to Auto or Man.

2. Set the sweep time for the Man mode.

Mode:
Range:

Resolution:

Auto, Man
22ms ~ 1000s (span>0Hz)
50us ~ 1000s (span=0Hz)
10us

Display Icon

The SWT icon will be displayed at the
bottom of the screen when in the sweep is
in manual mode.

Single Sweep

Description

The single sweep function is used to perform a
single sweep. When Sweep Single is pressed the
GSP-930 will perform a single sweep and then
stop.

Operation

1. Press

Sweep

> Sweep Single[F2] to put the

spectrum analyzer into single sweep mode.

2. Press Sweep Single[F2] again to perform a single

sweep.

Display Icon

The Sweep Single icon is displayed on the
right-hand side of the screen when the
sweep is in single mode.

Note

You must wait for the single sweep to finish before
pressing the Single Sweep key again.

If a setting is changed whilst the spectrum analyzer
is still sweeping, the single sweep will immediately
start over.