BASIC OPERATION

65

Average Type

Description

The Average Type function determines how the
GSP-930 determines the average value.

LOG Average: Averages the trace points on a
logarithmic scale.

Volt Average: Averages the amplitudes of the
trace points on a linear voltage scale.

Power Average: Averages the trace points on a
logarithmic scale in watts.

Operation

1. Press

BW/Avg

> Average Type[F5] and choose the

average type.

Range:

Default:

LOG Power, Volt
Average, Power Average
LOG Power