GW Instek GSP-930 User Manual User Manual

Page 166

background image

GSP-930 User Manual

164

Operation

1. Press

Limit

Line

>Pass/Fail Test.

2. To set a high limit, press High Limit[F1] and

choose one of the limit lines as the upper (high)
limit.

3. To set the low limit, press Low Limit[F2] and

select one of the limit lines as the lower limit.

4. Press Pass Criterion[F3] and select the pass

criteria.

Criteria:

All-In, Max-In, Min-In

5. Press Pass/Fail Test and turn the testing on.

6. The test result appears in the bottom of the

display.

Pass:

Fail:

Display Icon

The alarm icon is shown at the bottom of
the display whenever testing is turned on.

Note

At least one limit line (high or low) must be turned
on to enable testing.

If the high limit or low limit is turned off, the
maximum or minimum* display level is set
automatically as the high or low limit, respectively.

* +30dBm+Ref level offset or -120dBm+Ref level
offset