beautypg.com

Amplitude – GW Instek GSP-930 User Manual User Manual

Page 243

background image

APPENDIX

241

Phase Noise

Offset from
Carrier

Fc =1 GHz; RBW = 1
kHz, VBW = 10 Hz;
Average ≥ 40

10 kHz

<-88 dBc/Hz

Typical

2

100 kHz

<-95 dBc/Hz

Typical

1 MHz

<-113 dBc/Hz

Typical

Resolution Bandwidth (RBW) Filter

Filter Bandwidth 10 Hz to 3 kHz in 1-3-10

sequence

-3dB bandwidth
subtotal: 6 filters

10 kHz to 1 MHz,
increment in 10% step

-3dB bandwidth;
min. RBW = 10 kHz
@ zero span
Subtotal: 49 filters

200 Hz, 9 kHz, 120 kHz

-6dB bandwidth

Accuracy

± 8%, RBW ≥ 750 kHz

Nominal

3

± 5%, RBW < 750 kHz

Nominal

Shape Factor

< 4.5:1

Normal Bandwidth
ratio: -60dB:-3dB

Video Bandwidth (VBW) Filter

Filter Bandwidth 1 Hz to 1 MHz in 1-3-10

sequence

-3dB bandwidt

[1] Frequency Resolution = Span/(Sweep points - 1)
[2] Typical specifications in this datasheet mean that the performance can be
exhibited in 80% of the units with a 95% confidence level over the temperature
range 20 to 30 °C. They are not covered by the product warranty.
[3] Nominal values indicate expected performance. They are not covered by the
product warranty.

Amplitude

Amplitude Range

Measurement
Range

100 kHz to 1 MHz

Displayed Average
Noise Level (DANL)
to 18 dBm

1 MHz to 10 MHz

DANL to 21 dBm

10 MHz to 3 GHz

DANL to 30 dBm

Attenuator

Input Attenuator
Range

0 to 50 dB, in 1 dB step

Auto or manual
setup

Maximum Safe Input Level

Average Total
Power

≥ +33 dBm

Input attenuator
≥10 dB

DC Voltage

± 50 V