Continuous Sweep

Description

The GSP-930 has two main sweeping modes:
single and continuous. Use the continuous
mode to have the sweep constantly updated.

Operation

1. Press

Sweep

> Sweep Cont[F3] to put the

spectrum analyzer into continuous sweep
mode.

Display Icon

The Sweep Cont icon is displayed on the
right-hand side of the screen when the
sweep is in continuous mode.

Note

The GSP-930 will now continuously sweep unless
the mode is changed to single sweep mode or if
the system is waiting for a trigger condition.

Gated Sweep Overview

Description

The Gated Sweep mode allows a trigger signal
to dictate when the spectrum analyzer can
sweep. This mode is useful for characterizing
signals that are pulsed on and off, such as RF
burst transmissions or for measuring spurious
noise levels between transmission bursts.

Overview

1. The trigger signal must be synchronized to the

period of the input signal (shown as RF burst
below).

2. The start of the gate time is produced from the

positive or negative edge of the trigger signal +
the delay time.

3. The end of the gate time is determined by the