beautypg.com

Start and stop frequency – GW Instek GSP-930 User Manual User Manual

Page 42

background image

GSP-930 User Manual

40

Start and Stop Frequency

Description

The start/stop frequency function will set the
start and stop frequency of the span.

Operation

1. To set the start frequency, press

Frequency

> Start

Freq[F2] and enter the frequency and unit.

2. To set the stop frequency, press

Frequency

> Stop

Freq[F3] and enter the frequency and unit.

Range:
Resolution:
Default Start frequency:
Default Stop frequency:

0kHz~3GHz
1Hz
0Hz
3GHz

Display

Start Frequency

Start Frequency

Stop Frequency

Stop Frequency

Note

The start and stop frequency can change when the
span settings are used.
The stop frequency must be set higher than the
start frequency (for spans

0), otherwise the span

will be automatically set to 100Hz.