Vertical scale units, Reference level offset – GW Instek GSP-930 User Manual User Manual

Page 52

background image

GSP-930 User Manual

50

Vertical Scale Units

Description

Change the vertical units for both linear or
logarithmic scales.

Operation

1. Press

Amplitude

> More[F7]>Y Axis[F1] and then

choose the desired units.

2. The units are changed accordingly.

Units:

dBm, dBmV, dBuV,
Watts, Volts

Reference Level Offset

Description

The Reference Level Offset function sets an
offset value to the reference level to compensate
for any loss or gain from an external network or
device.
The offset value does not affect the input
attenuation or the on-screen trace.
This setting will change the reference level
readout, the scale readout and the marker
readout.

Operation

1. Press

Amplitude

> More[F7]>RefLvlOffset[F2] and set

the offset level and unit.

2. To remove the offset level, set the reference

offset to 0 dB.

Range:

0dB ~ 50dB

Display Icon

The AMP icon is displayed at the bottom
of the screen.