GSP-930 User Manual

54

Correction Table

5. Repeat steps 2 to 4 for any other correction

points.

6. To delete the selected point, press Delete

Point[F6].

7. Press Return[F7]>Save Correction[F5] to save the

correction set.

Note

Note that the correction points are automatically
sorted by frequency (low → high).
The correction set must be saved before it can be
turned on.

The frequency values

displayed

in the correction

table are rounded down for display purposes only.
The actual frequency for each point can be seen in
the Frequency soft-key.

Amplitude Correction On/Off

Description

Any one of the 5 correction sets can be turned
on.

Activate
Correction

1. Press

Amplitude

> More[F7]>Correction[F3]>

Correction Set[F1] and choose a correction set.

Correction Set:

1~5

2. Press Correction[F2] and toggle correction on.