FAQ

205

F

AQ

• I connected the signal but it does not appear on screen.
• I want to see which optional items are installed.
• The performance does not match the specification

I connected the signal but it does not appear on screen.

Run Autoset and let the GSP-930 find the best display scale for
your target signal. Press the Autoset key, then press Autoset[F1].
For details, see page 59.

I want to see which optional items are installed.

Check the optional items in the system information window. Press
the System key → System Information[F1]. For details, see page 105.

The performance does not match the specification.

Make sure the device is powered On for at least 30 minutes, within
+20°C~+30°C. This is necessary to stabilize the unit to match the
specification.

For more information, contact your local dealer or GWInstek at
www.gwinstek.com / [email protected].