beautypg.com

Yokogawa DAQWORX User Manual

Page 73

background image

4-26

IM WX102-01E

Absolute and relativeposition settings for DELTA, RRJC

When copying DELTA and RRJC channels, it is necessary to determine the relation to

reference channels.

• Absolute position

The reference channel of the copy destination doesn’t change.

For example, if the reference channel of Ch 005 is Ch 002, and you would copy Ch

005 settings to Ch 007, the reference channel of Ch 007 will remain Ch002.

However, if you are copying between units, the reference channel of the copy

destination will change.

For example, if the reference channel of Ch 005 is Ch 002, and Ch 002 settings are

TC, Type R, and you would copy Ch 005 settings to Ch 107, the reference channel of

Ch 107 will become Ch 102. Copying is, in this case, only possible if Ch 102, like Ch

002, is set to TC, Type R, and in addition, the scale must be set OFF.

• Relative position

The relative positions of channels and relative channels set in DELTA or RRJC, will

be maintained when copied. Therefore the reference channel of copy destination and

copy base will differ.

For example, if the reference channel of Ch 005 is Ch 002, and Ch 002 settings are

TC, Type R, and you would copy Ch 005 settings to Ch 009, the reference channel

of Ch 009 will become Ch 002. Copying is, in this case, only possible if Ch 009, like

Ch 002, is set to TC, Type R. It is advisable to let the copying include the reference

channel.