Icon Display during Monitoring/Recording
During Monitoring/Recording the following icons may be activated:
Hardware measurement condition icon
Math icon
Acquisition icon
Communication icon
Report condition icon
‘Communication’ icon
Plain icon :
Communication stop.
Icon with thunder mark :
Communicating.
Icon with exclamation ! mark : Communication error. Check cables and power supply.
‘Acquisition’ icon
Gray icon :
Acquisition stop.
Red blinking icon :
Acquiring data.
Red stable icon :
Acquisition stand by.
Icon with exclamation ! mark : Disk error. Saving space may be insufficient.
‘Math’ icon
Gray icon :
Not making calculations.
Icon with an X-mark on it :
No Math option available.
Blue blinking icon :
Calculating.
Blue stable icon :
Stopping Math.
Icon with exclamation ! mark : Math error. Hardware Math may not be processed in time.
‘Hardware measurement condition’ icon
Blue icon:
Normal measurement.
Red icon:
Hardware measurement ommision occurring.
‘Report condition’ icons
Grey icon :
Not making report
An x above an icon :
No report option available
Moving icon :
Making report
Stable icon :