Page 204: Error messages

11-2
IM WX102-01E
11.2 Error Messages and Corrective Actions
Displayed when incorrect settings have been entered or when the instrument is
malfunctioning. If the same problem still occurs after the proper corrective action has
been taken, diagnose the problem using the diagnosis software.
Warning Messages
No.
Message/Description
W1102
Converted data file will be overwritten. OK?
W3304
Start Initializing?
W3305
Start Reconstructing?
W3315
Stop Calibration?
W3316
Send Calibration Values?
W3317
60 mV, 200 mV have not been calibrated. Do you want to continue?
W3318
Store Calibration Values?
W3319
Abort Calibration?
W3320
Initial Balance Failed. Retry?
W3332
Close network?
W3333
Set the network address?
W3451
Delete the current Project?
W3671
Please change connected hardware to SET mode.
W3672
Hardware model doesn't match software setting. Continue sending?
W3673
Style numbers don't match. Continue sending?
W3674
Hardware and Software configurations don't match. Continue sending data?
W3675
Hardware and Software options don't match. Continue sending?
W3676
Some data couldn't be read.
W3677
Some existing data were changed.
W3678
This is a reference channel. Channels referring to this channel will be changed too!
W3679
Settings of other power module channels will be changed too!
W3680
Some channels out of the copy range will be changed too!
W3681
Some channels can't be copied.
W3682
Writing operation was changed.
W3683
Following the change of temperature unit, SET (Regular) settings have been initialized.
W3684
The digital print settings of some measurement and Math channels have been changed.
W3685
The relay settings of some measurement and math channels were set OFF.
W3686
Some Event/Action settings have been changed.
Error Messages
No.
Message/Corrective action
E0001
Insufficient Memory.
Exit other programs and restart, or reboot the OS and restart the program.
E0002
Insufficient Memory. Please close at once.
Exit other programs and restart, or reboot the OS and restart the program.
E0003
Can’t open shared memory.
Exit other programs and restart, or reboot the OS and restart the program.
E0004
Incorrect License number.
Reinstall with correct License number.
E0101
Please use DAQ32 launcher to open.
Please use DAQ32 Plus launcher to open.
Run the program from the launcher.
E0201
Can’t open. No YOKOGAWA binary file.
Cannot handle this file.
E0202
Can’t open. No DARWIN binary file.
Cannot handle this file.
E0203
No data available.
Cannot handle this file.
E0211
Can’t write to file.
Check the directory’s free space. Check that other programs are not using the file.