5 setting moving average and filter, 5 setting moving average and filter -14, Moving average settings – Yokogawa DAQWORX User Manual
Page 61: Filter settings

4-14
IM WX102-01E
4.5
Setting Moving Average and Filter
Moving Average Settings
1
To open the Channel Settings related setting screen, click the Channels tab, or
select Settings > SET (Regular) Settings > Channel Settings from the menubar.
2
Set the moving average in the pull-down menu which appears when you click the
setting area below the Mode label.
Moving Average (MOVE AVE)
Moving average can be set for each channel individually. The number of samples used
for the moving average can be set from 2 to 64. The default value is 0, which means
that no moving average is carried out. The result of the moving average is recorded and
displayed.
Filter Settings
1
To open the Channel Settings related setting screen, click the Channels tab, or
select Settings > SET (Regular) Settings > Channel Settings from the menubar.
2
Set the Filter ON or OFF in the pull-down menu which appears when you click the
setting area below the Filter label. Filter settings are only possible for pulse input.
Setting the Filter ON can eliminate pulse chattering of up to 5 ms.