External Trig In Setting

An external Low or High input signal can cause triggering. Under Setup select External 
Trig Setting
to choose External Trig In Setting: High Active, Low Active, or Toggle from
High to Low or Low to High once (3.3 V output).

Choose Port Speed

The default speed is Autospeed. You can also select the port speed from the 
drop‐down list: 1.5 Gbps, 3.0 Gbps, or 6.0 Gbps.

Note:

If a Port ID check box has no check, the analyzer does not capture any patterns for that port. The

system allocates trace memory for that port to its adjacent port, for example: I1, T1 <‐> I2, T2.

Ports Configuration

Select the Port Configuration button to display the Set Port Configuration dialog (see

Figure 2.74 on page 108

).

Figure 2.74: SAS: Set Port Configuration Dialog

The dialog shows the current port configuration. To select a port configuration, click the
down arrow to display the Select Port Configuration dialog .