Turn the K2 off and wait for a few seconds, then turn it back on.

The display should now show ELECRAFT for about two seconds,
followed by the frequency display. Now that the EEPROM is
initialized, this is the display you should always see on power-up. The
"R" and "T" in "ELECRAFT" appear in lower-case letters due to the
limitations of the seven-segment LCD characters.

Tap the

D I S P L AY

switch once to select voltage/current display.

The display should now show something similar to this:

E 1 2 .0 i0 .08


This would indicate that the power supply voltage (E) is about 12.0 V,
and the supply current (I) is about 80-100 mA.

7

Optical Encoder Test

Tap the

D I S P L AY

switch to return to the frequency display.

Turn the VFO knob in both directions and verify that the

displayed frequency changes accordingly.

Tap the

R AT E

switch to the right of the knob to change the

tuning rate, and repeat the VFO test at each rate.

Relay Test

Tap

B AN D +

. After a short delay the K2 will switch to the next

band. At the same time, you’ll hear one or more relays.

Tap the

B AN D +

switch 7 more times to verify that you hear

relays being switched with each band change. Note: The 1.8 MHz
(160 m) and 5 MHz (60 m) bands will not appear in the band list
unless the associated options are installed. This can be done only after
assembly and alignment have been completed.

7

The supply voltage reading will reflect a small drop across D10, the reverse-

polarity protection diode, typically 0.1 V on receive. Current readings may
vary at this stage of assembly, but should be under 150 mA.

Tap the

P R E / AT T N

switch three times. You should hear relays

switch each time.

RF Probe Assembly


The Switch Spacing Tool used during Front Panel can now be used as
the PC board for the RF probe. All parts for the probe, including a
ground alligator clip, 2 feet of RG174 coax, and banana plugs for a
DMM, are supplied with the kit. You can assemble the probe at any
time, using the instructions on page 9 of Appendix E.

Voltmeter Probe Assembly


If you do not have a DMM (digital multimeter), you can use the
simple DC voltage probe shown below in conjunction with the built-in
voltmeter. The crimp pin and 2-pin housing can be found in the
MISCELLANEOUS components bag.

Assemble the voltage probe as shown in Figure 6-13 using green-

insulated hookup wire. No ground connection is needed since you will
be measuring voltages inside the K2.

Hookup wire,

12” (0.3m)

Crimp pin

Housing

Tinned lead,
0.5” (13mm)

Figure 6-13

Plug the voltage probe assembly into P5 on the Control board.

The probe should be oriented so that the hookup wire is connected to
the (+) side of P5.