Device Input Tab

You can import analog (voltage) signal from up to two external devices and convert

it to digital NGC data via the signal import module (SIM).The SIM imports and

converts the signal to the appropriate trace units through its SIM channels. When

enabled, the details appear in the chromatogram’s legend, the trace appears on the

chromatogram, and the data can be analyzed in the Evaluation window. For each

device you can define its

Trace name

Minimum and maximum output range (in volts)

Trace unit type (displayed in the chromatogram’s legend)

Minimum and maximum unit range

To enable trace measurement from external detectors

1.

Ensure that the SIM is connected to the NGC instrument.

2.

On the back of the SIM, determine which external devices are connected to the
SIM 1 and SIM 2 ports.

3.

Select File > System Settings to open the System Settings dialog box.