The Maximizer mixer is for use with the Maximizer valve system. The Maximizer Mixer improves
gradient quality by mixing the output of the BioLogic DuoFlow workstation pumps. The mixer may be
used with one of two mixer barrels, which fit between the mixer body and the mixer top and are used
for higher flow rates. The mixer assembly comes pre-assembled with a 5 ml barrel extender, as shown
in Figure 1. This kit includes four screws (1/4-20 x 1/2”) to be used if the 5 ml mixer barrel is removed.
Refer to the following table when selecting a mixer volume.

Flow Rate

Barrel Extender

Mixer Capacity

Assembly Screws

0.5 to 10 ml/min

none

750 µl

1/4-20 x 1/2” (1.3 cm)

10 to 40 ml/min

5 ml Extender Barrel

5 ml

1/4-20 x 1 1/2” (3.8 cm)

40 to 80 ml/min

12 ml Extender Barrel

12 ml

1/4-20 x 3 1/4” (8.2 cm)

Figure 1. Maximizer Mixer

Note: Keep this literature with your DuoFlow Instruction Manual.

I. Connecting the Mixer to the BioLogic DuoFlow system.

a. Attach the mixer to a suitable position on a vertical or horizontal bar using its rod clamp. Attach the

mixer to the rack so that the outlet port faces up.

b. Connect the mixer signal cable (mini-DIN connector) to the connector marked “Mixer” on the rear

of the Maximizer.

c. Plug the unused input port using the plug provided for this purpose.

II. Adding or removing a mixer barrel.

Note: Prior to disassembly, flush any hazardous material and then drain the mixer. Use the
information in the Table above to select the appropriate mixer volume (the barrel) and screws.

MIXER

TOP

MIXER

BODY

12 ml MIXER

BARREL EXTENDER

OUTLET PORT

INLET PORT

5 ml MIXER

BARREL EXTENDER

MAXIMIZER MIXER

Maximizer Mixer

Assembly

Catalog Number 760-2010