Bio-Rad BioLogic Mixers User Manual
Maximizer mixer assembly
The Maximizer mixer is for use with the Maximizer valve system. The Maximizer Mixer improves
gradient quality by mixing the output of the BioLogic DuoFlow workstation pumps. The mixer may be
used with one of two mixer barrels, which fit between the mixer body and the mixer top and are used
for higher flow rates. The mixer assembly comes pre-assembled with a 5 ml barrel extender, as shown
in Figure 1. This kit includes four screws (1/4-20 x 1/2”) to be used if the 5 ml mixer barrel is removed.
Refer to the following table when selecting a mixer volume.
Flow Rate
Barrel Extender
Mixer Capacity
Assembly Screws
0.5 to 10 ml/min
none
750 µl
1/4-20 x 1/2” (1.3 cm)
10 to 40 ml/min
5 ml Extender Barrel
5 ml
1/4-20 x 1 1/2” (3.8 cm)
40 to 80 ml/min
12 ml Extender Barrel
12 ml
1/4-20 x 3 1/4” (8.2 cm)
Figure 1. Maximizer Mixer
Note: Keep this literature with your DuoFlow Instruction Manual.
I. Connecting the Mixer to the BioLogic DuoFlow system.
a. Attach the mixer to a suitable position on a vertical or horizontal bar using its rod clamp. Attach the
mixer to the rack so that the outlet port faces up.
b. Connect the mixer signal cable (mini-DIN connector) to the connector marked “Mixer” on the rear
of the Maximizer.
c. Plug the unused input port using the plug provided for this purpose.
II. Adding or removing a mixer barrel.
Note: Prior to disassembly, flush any hazardous material and then drain the mixer. Use the
information in the Table above to select the appropriate mixer volume (the barrel) and screws.
MIXER
TOP
MIXER
BODY
12 ml MIXER
BARREL EXTENDER
OUTLET PORT
INLET PORT
5 ml MIXER
BARREL EXTENDER
MAXIMIZER MIXER
Maximizer Mixer
Assembly
Catalog Number 760-2010