Bio-Rad ChromLab™ Software User Manual

Page 184

background image

6

| Creating a Method

182

| NGC Chromatography Systems and ChromLab Software

Column — test up to five column types for each column-switching valve in

use. This parameter requires at least one column-switching valve in the

fluidic scheme.

Note:

Pages 2 and 3 of the scouting wizard differ when scouting for

columns and samples. See

Using the Scouting Wizard to Scout Columns

on page

185

or

Using the Scouting Wizard to Scout Samples on page

187

for more information.

Sample — test up to seven different samples when one sample inlet valve

is used or up to 14 different samples when two sample inlet valves are

used (Port 8 is reserved for wash). This parameter requires at least one

sample inlet valve in the fluidic scheme.

Tip:

This option is available if, in the Sample Application phase, you

chose to load the loop through the sample pump or inject sample directly

onto the column. It is not available if you chose to load the loop manually.

Page 2: Select Method Steps to Scout

On this page you select the method steps to be scouted. The method steps that

include the parameter you selected on Page 1 of the scouting wizard are shown.