Running an experiment – Bio-Rad ChromLab™ Software User Manual
Page 105
![background image](/manuals/600362/105/background.png)
Running an Experiment
User Guide
| 103
Running an Experiment
Important:
When running an experiment manually, ensure that the valves are
set in the correct positions before starting the pump.
Clicking Start in the System Pump dialog box starts the run. You can change valve
positions while the pump is running and your manual run is in progress. You can
stop a run at any point and save it. You can then restart the pumps and continue the
run with a different name.
Note:
Complete the following procedures consecutively to perform a manual
run.
To select a fluidic scheme and install accessories
1.
(Optional) If your hardware configuration has changed, do one of the following:
Enter manual control mode on the touch screen.
On the computer, start ChromLab software and click Manual Run in the
Home window.
2.
Select Tools > Change Fluidic Scheme and choose the fluidic scheme that
matches the configuration of the devices connected to your NGC system.
Note:
If your NGC system includes multiple valves of the same type and
this is the first time you are using the fluidic scheme, the Fluidic Scheme
Mapping dialog box appears. Use this dialog box to map your NGC system
to the fluidic scheme.
3.
Install the accessories required to run the experiment:
a.
Connect a column.
b.
Install fraction collector racks and tubes.