beautypg.com

Bio-Rad ChromLab™ Software User Manual

Page 141

background image

Phase Controls and Parameters

User Guide

| 139

Loading the Loop with a Sample Pump

When a sample pump is used to load the loop, you can define the flow rate, the

volume to load into the loop, and, subsequently, the volume of sample to be

injected onto the column. You also have the option to interrupt loading the loop if air

is detected. If you select this checkbox and air is detected in any of the lines

connected to air sensors set to detect end of sample, the system stops the sample

pump and the method continues to the next step.

Fluidic Scheme Includes a Sample Inlet Valve

If your fluidic scheme includes at least one sample inlet valve and you choose to

load the loop through the sample pump, you have the option to perform a

pre-injection wash with buffer to flush the sample pump and lines and the loop. If

you select this checkbox, you can choose the port on the sample inlet valve from

which to draw buffer solution. You can define the flow rate and the volume of buffer

for the preinjection wash.

You also have the option to flush the sample pump and related connections after

sample injection. Again, you can choose the port on the sample inlet valve from

which to draw buffer solution and define the flow rate and the volume of buffer for

the postinjection wash.