Bio-Rad ChromLab™ Software User Manual
Page 39
Chromatogram View
User Guide
| 37
Table 1.
Trace
Definitions
Trace
Module
Explanation
UV
1–4
Single-Wavelength UV
and Multi-Wavelength
UV/Vis Detectors
UV — with Single-Wavelength UV module, a
reading of absorbance
MV — with Multi-Wavelength UV/Vis module, up
to four readings (
1–4)
Conductivity
Single-Wavelength UV
and Multi-Wavelength
UV/Vis Detectors
The conductivity of the fluid, read by the
conductivity monitor (mS/cm)
%B Theoretical
System Pump
Programmed by the user in the method or in the
System Pump dialog box
Gradient, system pump: % of pump B
Gradient, blending valve: % of valve Q4
Buffer blending: % of valve Q4
pH Theoretical
pH
Theoretical pH, as programmed in the method or
System Pump dialog box. Available only with the
buffer blending valve in buffer blending mode
pH
pH
The measure of pH read from the pH probe,
available only when a pH valve is present
System
Pressure
System Pump
Measured at the mixer (psi or MPa)
PreCol Pressure
Column Switching Valve
Available when at least one column switching
valve is present
Tip:
When two or more column switching
valves are present, the traces appear in
different shades of green on the
chromatogram. The trace number
corresponds to the valve number in the fluidic
scheme. For example, PreCol Pressure 1 in
the legend and the chromatogram
corresponds to C1 in the fluidic scheme.