| Evaluating Results
230
| NGC Chromatography Systems and ChromLab Software
Integrate and Delete
Integrate — detects peaks, using current parameter values, for the selected traces
and updates the Peaks table with detected values.
Delete — deletes previously calculated peaks. All peaks are removed only from
traces selected in the traces list.
Starting Peak Integration
When you select Peak Integration, a single run becomes an analysis, and the
system prompts you for an analysis name when you first save it.
To start peak integration for the first time
Click Peak Integration on the Run tab toolbar.
The Peak Integration pane opens, and peak integration is performed
automatically using default settings.
To perform peak integration with changed settings
Select the traces of interest, change the settings in the Peak Integration pane,
and then click Integrate at the bottom of the pane.
Updated peak integration results appear in the chromatogram and in the Peaks
table.
Tip:
You can apply different settings for individual traces by integrating each
trace separately with its own settings. Clear the checkbox for the other traces
in the list and click Integrate.