Performing Multicolumn Purification Chromatography

User Guide

| 291

Performing Multicolumn Purification Chromatography

To perform multicolumn purification chromatography on an NGC system you must

Plumb the NGC system

Prime the NGC system

Select and save a Method template

Run the method

Plumbing the NGC System

How you set up and plumb the NGC system depends on the type of method you

plan to run. This section provides information about plumbing the system to use the

multicolumn purification templates.

Plumbing NGC Discover Pro Systems

This table lists the tasks for plumbing NGC Discover™ Pro systems. See Chapter 5,

Plumbing and Priming the NGC Chromatography Systems, in the NGC

Chromatography Software Installation Guide for detailed instructions.

Plumbing NGC Discover Pro Systems

Task

1.

Tube #6 between the outlet port on the sample inlet valve and the
inlet port on the sample pump

2.

Connect one of the following:

Tube #4 from the Common inlet port on the BioFrac fraction
collector diverter valve to port 1 on the outlet valve

Waste tubing to port 1 on the outlet valve

3.

Inlet tubing to the remaining ports on the sample inlet valve

4.

Outlet tubing to the remaining ports on the outlet valve

Note:

Remove the inlet tubing from the sample pump inlet

manifold if it is installed.