Bio-Rad ChromLab™ Software User Manual

Page 190

background image

6

| Creating a Method

188

| NGC Chromatography Systems and ChromLab Software

include a preinjection sample pump wash, the method is updated and you

proceed to Page 3. If you click No in the message box (if, for example, you are

using the same sample), the method is not updated and you proceed to Page

3.

Page 3:

On this page you can name each run and include or exclude a sample in the scout

sequence. The scout sequence table comprises the following columns:

Run # — the scout run number. This field is not editable.

Run Name — the scout run name. The autogenerated name is based on

the sample inlet valve module number and the port number on that valve.

This field is editable.

Tip:

To edit the run name column easily, create a list in Microsoft Word or

Excel. Copy and paste the list into the column to replace the

auto-generated names.

Sample Port Position — this field is not editable.

Injection Volume (ml) —this field is not editable.