How to set the id no – Dwyer UXF2 User Manual
Page 99
- 89 -
4.7.9. How to set the ID No.
Description
z
Set the ID No. for protection of parameters (Section 4.4.1.).
If ID No. is set, the number must be inputted before canceling the parameter protection.
z
To validate the parameter protection, set the parameter protection to "PROTECTION ON" (see Section 4.4.1.).
ID No. settable range: 0000 to 9999 (4 digits).
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).
Typical operation Set the ID No. to "1106".
Keying order
Description
Indication
ź
Pressed 4 times to indicate "MAINTENANCE MODE".
MAINTENANCE MODE
ENT
ź
Pressed to indicate "CURRENT CALIBRATION".
CURRENT
CALIBRATION
ź
Pressed 8 times to indicate "ID NO."
REGISTER ID NO.
ENT
ź
Pressed 2 times to blink the 2nd line.
REGISTER ID NO.
0
000
ź
Press
key and
key for composing "1106".
REGISTER ID NO.
110
6
Pressed to register it.
REGISTER ID NO.
½½ COMPLETE ½½
Ļ
ENT
ź
ź
ź
ź
--- ID No. registered.---
REGISTER ID NO.
½½½½
ESC
Using
ESC
key and
key, resume the
measurement mode.
Note: To validate the parameter protection, set the
parameter protection to "PROTECTION ON".