2. SELECTION OF INSTALLATION PLACE

Select an installation place taking into account the following factors from the viewpoint of ease of maintenance and
checkup, instrument life, and reliability.

(1) A place where ambient temperature and humidity are –20 to +50

qC and 90% RH or less for flow

transmitter (UXF2), and –20 to +60

qC and 90% RH or less for detector SX3.

(2) A place not exposed to direct sunshine nor effects of weather.
(3) Sufficient space for daily inspection, wiring, etc.
(4) A location not subject to radiated heat from a heating furnace, etc.
(5) A location not subjected to corrosive atmosphere.
(6) A location not subject to flooding.
(7) A location free from excessive vibration, dust, dirt and moisture.

2.1. Flow transmitter

Secure at least 100 mm of space between the flow
transmitter and nearby wall. Also secure a space for
opening the front cover for maintenance.
Secure a cable wiring space under the enclosure.

O P

E

N

133

100

133

370 min.

min.

Top view of mounting