Typical calculation

Calculate the range that permits setup of the pulse value under the range (FLOW SPAN-1) and pulse width indicated
below.
When set values of the range (FLOW SPAN-1) and pulse width are as follows;

FLOW SPAN-1: 36 [m

3

/h] (=0.01 [m

3

/s]), PULSE WIDTH: 50 [ms]

i) Case of DO1 output

From condition 1,

FLOW SPAN-1 [m

3

/s]

0.01 [m

3

/s]

PULSE VALUE

100 [Hz]

=

100 [Hz]

PULSE VALUE

FLOW SPAN-1[m

3

/s]

u 86400 [s] = 0.01 [m

3

/s]

u 86400 [s] = 864 [m

3

]

From the above:

0.1 [L]

PULSE VALUE 864 [m

3

] ····················· A

From condition 2,

2

u PULSE WIDTH [ms]

2

u 50 [ms]

PULSE VALUE

FLOW SPAN-1 [m

3

/s]

u

1000

= 0.01 [m

3

/s]

u

1000

= 0.001 [m

3

] = 1 [L] ·········································· B

The settable range of the pulse value that satisfies both condition 1 and condition 2 is as follows from results
of calculation A and B:

1 [L]

PULSE VALUE 864 [m

3

]

ii) Case of DO2 output

From condition 1,

FLOW SPAN-1 [m

3

/s]

0.01 [m

3

/s]

PULSE VALUE

1 [Hz]

=

1 [Hz]

Condition 2 is same as that of the case of DO1 output indicated above.

Therefore, the settable range of the pulse value is as follows from results of calculation B and C:

10 [L]

PULSE VALUE 864 [m

3

]

For actual keying, refer to the typical operation indicated in the next page. Set the parameter protection to OFF
beforehand (Section 4.4.1).

= 0.0001 [m

3

] = 0.1[L]

= 0.01 [m

3

] = 10 [L] ······················ C