Output setup mode, Adjusting zero point – Dwyer UXF2 User Manual
Page 56
- 46 -
4.5. Output setup mode
4.5.1. Adjusting zero point
Description
z
Zero point is calibrated.
Settable range
CLEAR:
Clears the zero point calibration value to "0". Used in case the flow cannot be stopped when
calibrating the zero point.
Note: 1. When possible, stop the flow and carry out "SET ZERO" stated below. Otherwise, an
error may occur in the zero point.
SET ZERO: A point where "SET ZERO" is carried out is regarded as zero. Used for calibrating the zero point
upon stopping the flow.
Note: 2. The flow must completely be stopped. Otherwise, the flowing status is regarded as zero,
thereby causing an error.
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).
Typical operation Completely fill the piping, close the upstream and downstream valves, and proceed to zero point
calibration.
Keying order
Description
Indication
ź
Pressed twice to indicate "OUTPUT SETUP".
OUTPUT SETUP
ENT
ź
Pressed twice to indicate "ZERO ADJUST" and blink the
cursor.
ZERO ADJUST
CLEAR
ź
Pressed to select "SET ZERO".
ZERO ADJUST
SET ZERO
Pressed to execute "SET ZERO".
* Be sure to completely stop the flow beforehand.
ZERO ADJUST
½½ COMPLETE ½½
Ļ
ENT
ź
ź
ź
ź
--- Zero point calibrated.---
ZERO ADJUST
SET ZERO
ESC
Press
ESC
key once, and
key 3 times to resume
the measurement mode.
0.000 m/s
0.000 m3/h
BULLETIN F-70