4.5. Output setup mode

4.5.1. Adjusting zero point

Description
z

Zero point is calibrated.

Settable range

CLEAR:

Clears the zero point calibration value to "0". Used in case the flow cannot be stopped when
calibrating the zero point.
Note: 1. When possible, stop the flow and carry out "SET ZERO" stated below. Otherwise, an

error may occur in the zero point.

SET ZERO: A point where "SET ZERO" is carried out is regarded as zero. Used for calibrating the zero point

upon stopping the flow.
Note: 2. The flow must completely be stopped. Otherwise, the flowing status is regarded as zero,

thereby causing an error.

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).

Typical operation Completely fill the piping, close the upstream and downstream valves, and proceed to zero point

calibration.

Keying order

Description

Indication

ź

Pressed twice to indicate "OUTPUT SETUP".

OUTPUT SETUP

ENT

ź

Pressed twice to indicate "ZERO ADJUST" and blink the
cursor.

ZERO ADJUST

CLEAR

ź

Pressed to select "SET ZERO".

ZERO ADJUST

SET ZERO

Pressed to execute "SET ZERO".
* Be sure to completely stop the flow beforehand.

ZERO ADJUST

½½ COMPLETE ½½

Ļ

ENT

ź
ź
ź
ź

--- Zero point calibrated.---

ZERO ADJUST
SET ZERO

ESC

Press

ESC

key once, and

key 3 times to resume

the measurement mode.

0.000 m/s
0.000 m3/h

BULLETIN F-70