5.7. Troubleshooting
5.7.1. If indication is abnormal
Symptom Cause
No indication (blank).
z
Power is not turned on.
z
Source voltage is too low.
z
Fuse is blown.
z
LCD is defective.
To “Section 5.7.6. Remedying a hardware
fault.”
z
ˠ˟ power polarity is inverted.
1st line is indicated black.
z
Source voltage is too low.
z
DC power polarity is inverted.
z
LCD is defective.
To “Section 5.7.6. Remedying a hardware
fault.”
Indication is undefined.
z
Hardware error.
To “Section 5.7.6. Remedying a hardware
fault.”
Indication is dim.
z
Ambient temperature is too low (below -20
qC).
Raise the temperature.
z
End of LCD unit life.
Replace the LCD unit.
Entirely black
z
Ambient temperature is too high (50
qC or higher).
Lower the temperature.
Some characters on the LED are
missing.
LED is extinguished.
z
Check the LCD/LED, following the procedure in “5.7.1.1. Checking
the LCD/LED.” Missing dots on LCD. LED is not lit.
To “Section 5.7.6. Remedying a hardware fault.”
LED is lit in red.
z
Wave receiving error
Check by following the procedure in
“5.7.1.2. Checking the LED lit in red.”
5.7.1.1. Checking the LCD/LED
Follow the procedure below to check possible display errors.
Keying order
Description
Indication
ź
Press 4 times to display “MAINTENANCE MODE.”
MAINTENANCE MODE
ENT
ź
Press once to display “CURRENT CALIBRATION.”
CURRENT
CALIBRATION
ź
Press 11 times to display “Check LCD/LED.”
LCD/LED CHECK
ENT
ź
Press once.
ź
Each time the key is pressed, the display is switched in
the order shown below.
LCD: OFF completely
LED: Lit in green
LCD: Darkened
LED: Lit in red
If dots on the LCD are missing or the LED does not come
on, the LCD/LED may have failed.
● Lit in red
● Lit in green