3.9. Setting the range and total pulse output

The range is set by the following procedure.
According to a selected range value, an analog output (4-20 mA DC) is set.
A pulse is outputted every time the integrated value reaches a pulse value.
(Note: Must be completed after setting the piping parameters in Section 3.4.)

(1) Selecting a unit system: Metric or inch system

Note: Factory set at "Metric system". For metric units, skip to step (2) below.
Proceed to the following menu from the measurement mode.

Keying LCD

indication/comment

key pressed 3 times.

1st line: [MEASURE SETUP].

ENT

key pressed.

1st line: [SYSTEM UNIT]. 2nd line: [METRIC]. * As selected currently.

ENT

key pressed.

2nd line blinks.

key pressed.

Select a unit system out of metric system: [METRIC] and inch system:
[ENGLISH].

ENT

key pressed.

Registered after [**COMPLETE**] is displayed about 1 sec on 2nd line.

(2) Selecting a flow rate unit: L/s, m3/h or other flow rate unit.

Follows the operation from (1) above.

Keying LCD

indication/comment

key pressed 1 times.

1st line: [FLOW UNIT], 2nd line: [L/s] * As selected currently.

ENT

key pressed.

2nd line blinks.

key pressed.

Repeatedly until a desired flow rate unit is selected.

ENT

key pressed.

Registered after [**COMPLETE**] is displayed about 1 sec on 2nd line.

(3) Selecting a total unit: mL, L, m3, or other total unit.

* Must be selected when total indication or total pulse output is used.
Follows the operation from (2) above.

Keying LCD

indication/comment

key pressed 1 times.

1st line: [TOTAL UNIT], 2nd line: [mL] * As selected currently.

ENT

key pressed.

2nd line blinks.

key pressed.

Repeatedly until a desired total unit is selected.

ENT

key pressed.

Registered after [**COMPLETE**] is displayed about 1 sec on 2nd line.

ESC

key pressed.

1st line: [MEASURE SETUP]

key pressed twice.

Resumes the measurement mode.

* Carrying out the operation in (1) to (3) above completes setting of the unit system, flow rate unit and total unit.