Dwyer UXF2 User Manual
Page 112
- 102 -
Symptom Cause
Remedy
Measurement value
contains an error.
z
Inputted pipe specifications are
different from actual values.
z
Scales exist on wall of old pipe.
Difference in inner diameter of 1%
produces about 3% of error.
z
Input correctly.
z
When inputting, regard the scales
as lining.
z
Straight pipe is inadequate
(10D upstream and 5D
downstream cannot be secured).
Find a better place for mounting the
sensor (upstream the disturbance).
There must be no disturbance within
30D upstream.
Pump, valve, joint pipe, etc. are not
allowed.
z
Mount the sensor at different
angles with respect to pipe cross-
section until an average is
obtained.
z
Pipe is not filled with water or
contains mud or sand.
Reading rises as cross-sectional area
reduces.
z
Move to vertical pipe.
BULLETIN F-70
- 166T (1 page)
- 160S (2 pages)
- PAFS-1000 (1 page)
- 160F (2 pages)
- Mark II WSI (4 pages)
- MW-1 (2 pages)
- 89088 (12 pages)
- 660 (2 pages)
- 670 (16 pages)
- FLST (4 pages)
- AFG (6 pages)
- FAFM (2 pages)
- DAFM (2 pages)
- STRA (4 pages)
- 530 (2 pages)
- AVFS (2 pages)
- 641RM (4 pages)
- 641B (4 pages)
- 641 (4 pages)
- VTT (2 pages)
- SFI-700 (2 pages)
- GFM (32 pages)
- GFC (40 pages)
- DMF (44 pages)
- GFM2 (20 pages)
- GFM3 (22 pages)
- TFP-LI (30 pages)
- TFP-GI (27 pages)
- TBS (7 pages)
- HFT (4 pages)
- CAM (6 pages)
- PWD (3 pages)
- BAT (3 pages)
- FPCA (4 pages)
- FPCA (2 pages)
- SPK (4 pages)
- MFS (8 pages)
- LOFM (3 pages)
- FLMG (12 pages)
- MFS2 (16 pages)
- PFT (4 pages)
- IPFS (7 pages)
- UFB (48 pages)
- UFC (56 pages)
- Series PUB (44 pages)