background image

- 79 -

4.7. Maintenance mode

4.7.1. How to calibrate the analog output

Description
z

The calibration is performed to output 4 mA and 20 mA when the analog signal (4-20 mA DC) output is 0%
and 100%, respectively.

z

Connect an ammeter to Iout terminals as shown below. In the CURRENT CALIBRATION mode, select 4 mA or

20 mA, and operate

key (UP) or

key (Down).

+

-

+

Iout

Ammeter

+

DO1

+

DO2

-

Iout

-

DO1

-

DO2

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).

Typical operation Calibrate the output of 4 mA and 20 mA.

Keying order

Description

Indication

ź

Pressed 4 times to indicate "MAINTENANCE MODE".

MAINTENANCE MODE

ENT

ź

Pressed to indicate "CURRENT CALIBRATION".

CURRENT
CALIBRATION

ENT

ź

Pressed twice to select the calibration mode for 4 mA.
(When the cursor blinks, 4 mA of current is outputted.)

Then, press

(UP) or

(Down) key so as to

obtain 4 mA, watching the reading on ammeter or other
calibration instruments.

CURRENT

4 mA

Pressed to register the calibration result.

CURRENT

½½ COMPLETE ½½

Ļ

ENT

ź
ź
ź
ź

--- 4 mA calibration registered.---

CURRENT
4 mA

ź

Pressed to select 20 mA.

CURRENT
20 mA

ENT

ź

Pressed to select the calibration mode for 20 mA.
(When the cursor blinks, 20 mA of current is outputted.)

Then, press

(UP) or

(Down) key so as to

obtain 20 mA.

CURRENT

20 mA

Pressed to register the calibration result.

CURRENT

½½ COMPLETE ½½

Ļ

ENT

ź
ź
ź
ź

--- 20 mA calibration registered.---

CURRENT
20 mA

ESC

Press

ESC

key twice, and

key once to resume

the measurement mode.

0.000 m/s
0.000 m3/h

BULLETIN F-70