Dwyer UXF2 User Manual

Page 109

background image

- 99 -

5.7.1.3. Displaying the data in maintenance mode

Follow the procedure shown below to check possible display errors.

Keying order

Description

Indication

ź

Press 4 times to display “MAINTENANCE MODE.”

MAINTENANCE MODE

ENT

ź

Press once to display “CURRENT CARIBRATION.”

CURRENT

CALIBRATION

ź

Press 12 times to display “DATA DISPLAY.”

DATA DISPLAY

ENT

ź

Press once.

• Displays the measurement value of transit time, T1

(forward time), and T2 (reverse time).

T1: 0.0000

usec

T2: 0.0000

usec

ź

Press once.

• Displays the measurement value of average transit

time, T0, and transit time difference, DT.

T0: 0.0000

usec

DT: 0.00 nsec

ź

Press once.

Displays the calculated value of pass time of the

substances other than fluid, Ta, and angle of
incidence of the fluid,

T.

Ta: 0.0000

usec

LJ: 0.000°

ź

Press once.

• Displays the calculation value of sound velocity in

fluid, Cf, and Reynolds number, Re.

Cf: 0.0 m/s
Re: 0

ź

Press once.

• Displays correction coefficient of flow velocity

distribution, K, and flow velocity, V.

K: 1.3333
V: 0.000 m/s

ź

Press once.

• Displays the intensity of received signals. The

smaller the value, the larger the intensity of received
signals. Normal measurement values fall within the
range from 4700 to 8000. If the display appears as
8484, no signals are being received. Ultrasonic
waves may not be transmitted because of insufficient
water volume or rust of piping.

AGC U:

08484

AGC D:

08484

ź

Press once.

• Displays the peak value of received signal waveform.

Normal values stably fall within the range from 395
to 415. If the value fluctuates significantly, objects
that constitute barriers against ultrasonic wave
transmission such as air bubbles or foreign matter
may be contained in the fluid. Stop the flow and
check if normal value is resumed. If so, there is a
possibility that air bubbles are contained.

P/H U:

164

P/H D:

164

ź

Press once.

• Displays the state of the ultrasonic transmission

voltage circuit. It is normal if the value for DRV_R
falls within the range from 810 to 890.

DRV S:

227

DRV R:

845

ESC

Obtain a measurement-mode display using the

ESC

and

the

keys.

0.000

m/s

0.000 m3/h

BULLETIN F-70