Setting the lcd indication – Dwyer UXF2 User Manual
Page 59
- 49 -
4.5.4. Setting the LCD indication
Description
z
Flow velocity indication
Selectable flow velocity units : m/s (if SYSTEM UNIT was set to METRIC)
: ft/s (if SYSTEM UNIT was set to ENGLISH) (Section 4.6.1).
Note: 1. The decimal point position is fixed.
z
Flow rate indication
Selectable flow rate indications: Actual value reading, % reading.
Note: 2. The indication unit is as selected by FLOW UNIT (Section 4.6.2).
z
Total indication
Selectable total indications: Actual total value reading (forward/reverse flow), total pulse count (forward/reverse
flow).
Note: 3. When total unit is changed, perform “RESET”. (Refer to item 4.5.6)
Conversion of totalizer memory contents due to unit change is not automatic.
z
How to verify the indication
Set the DISPLAY setting mode to 1st. ROW (for indication on 1st line) or 2nd. ROW (for indication on 2nd line),
and then select indication contents.
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).
Typical operation Set the LCD indication on 1st line to % reading.
Keying order
Description
Indication
ź
Pressed twice to indicate "OUTPUT SETUP".
OUTPUT SETUP
ENT
ź
Pressed to indicate "ZERO ADJUST".
ZERO ADJUST
CLEAR
ź
Pressed 3 times to indicate "DISPLAY".
DISPLAY
1ST. ROW
ENT
ź
Pressed to blink the cursor.
DISPLAY
1ST. ROW
ENT
ź
Pressed again to select "1st. ROW".
1ST. ROW
VELOCITY
ź
Pressed twice to select "FLOW RATE (%)".
1ST. ROW
FLOW RATE (%)
ENT
ź
Pressed to enter "FLOW RATE (%)" and indicate "1st.
ROW DIGIT".
1ST. ROW DIGIT
½½½½. ½½
ź
Pressed to shift the decimal point position to next place.
1ST. ROW DIGIT
½½½½½. ½
Pressed to register it.
1ST. ROW DIGIT
½½ COMPLETE ½½
Ļ
ENT
ź
ź
ź
ź
--- FLOW RATE (%) indication validated.---
1ST. ROW
FLOW RATE (%)
ESC
Press
ESC
key 2 times, and
key 3 times to
resume the measurement mode
0.0 %
0.000 m3/h
BULLETIN F-70