beautypg.com

Checking the led lit in red, Bulletin f-70 – Dwyer UXF2 User Manual

Page 108

background image

- 98 -

ESC

Obtain a measurement-mode display using the

ESC

and the

keys.

0.000

m/s

0.000 m3/h

5.7.1.2. Checking the LED lit in red

Check the LED lit in red, following the procedure shown below.

LED lit in red

Check the AGC_U and AGC_D

values.

The value appears as
8484.

Check P/H_U and P/H_D.

The value fluctuates.

The value is kept
stable between 395
and 415.

Check T1 and T2.

Both AGC_U and AGC_D

1

ɦ Check AGC_U, AGC_D, P/H_U, P/H_D, T1 and T2, following the procedure in “5.7.1.3

Displaying the data in maintenance mode.”

Error status

: No waves are received.

Cause and remedy :

Transmission error due to scale or rust within piping

Ÿ Move the mounting position of the detector.

Break of signal cable

Ÿ Repair or replace the signal cable.

Detector mounting failure

Ÿ Remount the detector.

Transmitter circuit failure or detector failure

Ÿ See “5.7.6

Remedying a hardware fault

.”

Error status

: No waves are received.

Cause and remedy :

Transmitter circuit failure (one-side receiving circuit failure)

Ÿ See “5.7.6

Remedying a hardware fault

.”

Either AGC_U or AGC_D

Error status

: Received waves fluctuate.

Cause and remedy :

Mixing in of air bubbles

Ÿ Stop the flow, and if the measurement is found to be normal,
there is great possibility that air bubbles are entrained. Remove
bubbles in such cases.

Noise

Ÿ Check ultrasonic signals for existence of noise using an
oscilloscope. Remove noise if found by performing grounding,
etc.

1

Error status

: Noise

Cause and remedy :

Ÿ Check ultrasonic signals for existence of noise using an
oscilloscope. Remove noise if found by performing
grounding, etc.

Error status

: Operation error

Cause and remedy :

CPU operation error

Ÿ Reset the power (OFFoON)
If the power is not restored, see “5.7.6

Remedying

a hardware fault

.”

T1 and T2 fluctuate
or differ significantly
from each other.

YES

YES

NO

YES

NO

YES

NO

YES

NO

NO

BULLETIN F-70