beautypg.com

Bulletin f-70 – Dwyer UXF2 User Manual

Page 27

background image

- 17 -

Keying

LCD indication/comment

and

key pressed.

Input the wall thickness of the measurement pipe. As necessary, check the piping data
in section 6.6.

: Selects a numeric.

: Shifts the place.

ENT

key pressed to enter.

Registered after [**COMPLETE**] is indicated about 1 second on 2nd line.

key pressed.

1st line: [LINING MATERIAL]. 2nd line: [NO LINING]. * As selected currently.

If pipe is not lined, press

key to go to selection of next fluid to be measured.

ENT

key pressed.

Cursor blinks on 2nd line.

key pressed to select.

Select the lining material from menus. If there is no corresponding menu, input the
sound velocity of lining material on sound velocity input screen.
If necessary, see lining data in section 6.6.

ENT

key pressed to enter.

Registered after [**COMPLETE**] is indicated about 1 second on 2nd line.

key pressed.

1st line: [LINING THICKNESS]. 2nd line: [2.00 mm]. * As selected currently.
Note: Not indicated if lining material is set at [NO LINING].

ENT

key pressed.

Cursor blinks on 2nd line.

and

key pressed.

Input the lining thickness.

: Selects a numeric.

: Shifts the place.

ENT

key pressed to enter.

Registered after [**COMPLETE**] is indicated about 1 second on 2nd line.

key pressed.

1st line: [KIND OF FLUID]. 2nd line: [WATER]. * As selected currently.

ENT

key pressed.

Cursor blinks on 2nd line.

key pressed to select.

Select [WATER] or [SEA WATER]. In case of other fluid, input the sound velocity of
fluid on sound velocity input screen.
If necessary, see piping data in section 6.6.

ENT

key pressed to enter.

Registered after [**COMPLETE**] is indicated about 1 second on 2nd line.

key pressed.

1st line: [KINEMATIC VISCO]. 2nd line: [1.0038E–6m2/s]. * As selected
currently. Kinematic viscosity of water is factory set. If fluid to be measured is
other than water, input the kinematic viscosity referring to piping data in section 6.6.

ENT

key pressed.

Cursor blinks on 2nd line.

and

key pressed.

Input the kinematic viscosity.

: Selects a numeric.

: Shifts the place.

ENT

key pressed to enter.

Registered after [**COMPLETE**] is indicated about 1 second on 2nd line.

key pressed twice.

1st line: [SENSOR TYPE]. 2nd line: [SX3-A0/B0]. * As selected currently.

ENT

key pressed.

Cursor blinks on 2nd line.

key pressed.

Select [SX3-A0/B0] or [SX3-C0/D0].

ENT

key pressed to enter.

Registered after [**COMPLETE**] is indicated about 1 second on 2nd line.

ESC

key pressed.

1st line: [PIPE PARAMETER]. 2nd line: [S= 16 ( 48mm)] in case SX3-A/B

Note Sensor unit spacing calculated by above settings is indicated for sensor unit

spacing at detector installation.

ESC

key pressed.

1st line: [MEASURE SETUP]

key pressed twice.

Measurement mode is resumed.

BULLETIN F-70