How to set the constant current output – Dwyer UXF2 User Manual

Page 90

background image

- 80 -

4.7.2. How to set the constant current output

Description
z

Generates a fixed value output of analog signal.
Application example: The operation of a connected ammeter is checked by generating a fixed value output of analog
signal.

z

In the constant current setting mode (CURRENT OUTPUT SETTING), set the constant current output value.
Settable range: -20% (0.8 mA) to +120% (23.2 mA).

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).

Typical operation Set the constant current output of 50% (12 mA).

Keying order

Description

Indication

ź

Pressed 4 times to indicate "MAINTENANCE MODE".

MAINTENANCE MODE

ENT

ź

Pressed to indicate "CURRENT CALIBRATION".

CURRENT
CALIBRATION

ź

Pressed to indicate "CURRENT OUTPUT SETTING".

CURRENT
OUTPUT SETTING

ENT

ź

Pressed to select the setting screen.

CURRENT SETTING
4.0 mA

ENT

ź

Pressed to blink the cursor.
Note: A constant current output starts.

CURRENT SETTING

0

4.0 mA

ź

Operated to compose [12].

CURRENT SETTING
1

2

.0 mA

Pressed to output 12 mA.

CURRENT SETTING

½½ COMPLETE ½½

Ļ

ENT

ź
ź
ź
ź

--- 12 mA is being outputted.---

CURRENT SETTING
12.0 mA

ESC

ź

Pressed to stop the constant current output.
Note: The current output at a measurement status is

resumed.

CURRENT
OUTPUT SETTING

ESC

Press

ESC

key and

key to resume the

measurement mode.

0.000 m/s
0.000 m3/h

BULLETIN F-70