How to set the constant current output – Dwyer UXF2 User Manual
Page 90
- 80 -
4.7.2. How to set the constant current output
Description
z
Generates a fixed value output of analog signal.
Application example: The operation of a connected ammeter is checked by generating a fixed value output of analog
signal.
z
In the constant current setting mode (CURRENT OUTPUT SETTING), set the constant current output value.
Settable range: -20% (0.8 mA) to +120% (23.2 mA).
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).
Typical operation Set the constant current output of 50% (12 mA).
Keying order
Description
Indication
ź
Pressed 4 times to indicate "MAINTENANCE MODE".
MAINTENANCE MODE
ENT
ź
Pressed to indicate "CURRENT CALIBRATION".
CURRENT
CALIBRATION
ź
Pressed to indicate "CURRENT OUTPUT SETTING".
CURRENT
OUTPUT SETTING
ENT
ź
Pressed to select the setting screen.
CURRENT SETTING
4.0 mA
ENT
ź
Pressed to blink the cursor.
Note: A constant current output starts.
CURRENT SETTING
0
4.0 mA
ź
Operated to compose [12].
CURRENT SETTING
1
2
.0 mA
Pressed to output 12 mA.
CURRENT SETTING
½½ COMPLETE ½½
Ļ
ENT
ź
ź
ź
ź
--- 12 mA is being outputted.---
CURRENT SETTING
12.0 mA
ESC
ź
Pressed to stop the constant current output.
Note: The current output at a measurement status is
resumed.
CURRENT
OUTPUT SETTING
ESC
Press
ESC
key and
key to resume the
measurement mode.
0.000 m/s
0.000 m3/h
BULLETIN F-70