4.5.6.2. Setting the preset value

Description
z

Preset value: Value which appears on the total counter when the total value has been reset.
Settable range: 0 to 99999999

Notes

1. A resetting action simultaneously resets both forward total memory and reverse total memory.
2. Set the total unit beforehand in the MEASURE SETUP mode (see Section 4.6.3.).

Total value

Preset value

Time

Reset

Reset

Reset

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1)

Typical operation Set the preset value to 100 m

3

.

* Set the total unit beforehand.

Keying order

Description

Indication

ź

Pressed twice to indicate "OUTPUT SETUP".

OUTPUT SETUP

ENT

ź

Pressed to indicate "ZERO ADJUST".

ZERO ADJUST
CLEAR

ź

Pressed 4 times to indicate "RANGE".

RANGE
FLOW RATE

ENT

ź

Pressed to blink the cursor.

RANGE

FLOW RATE

ź

Pressed to indicate "TOTAL" on 2nd line.

RANGE

TOTAL

ENT

ź

Pressed to indicate "TOTAL MODE".

TOTAL MODE
START

ź

Pressed 3 times to indicate "TOTAL PRESET".

TOTAL PRESET
0 m3

ENT

ź

Pressed to indicate the cursor.

TOTAL PRESET

0

00000000 m3

ź

Pressed 6 times to move the cursor.
* Note that input cannot be entered on the firsr digit (the

leftmost digit).

TOTAL PRESET
000000

0

00 m3

ź

Pressed to select "1".

TOTAL PRESET
000000

1

00 m3

Pressed to register it.

TOTAL PRESET

½½ COMPLETE ½½

Ļ

ENT

ź
ź
ź
ź

--- PRESET value registered.---

TOTAL PRESET
100 m3

ESC

Press

ESC

key 3 times, and

key 3 times to

resume the measurement mode.

0.000 m/s

0.000 m3/h

BULLETIN F-70