Bulletin f-70 – Dwyer UXF2 User Manual
Page 70
- 60 -
4.5.6.2. Setting the preset value
Description
z
Preset value: Value which appears on the total counter when the total value has been reset.
Settable range: 0 to 99999999
Notes
1. A resetting action simultaneously resets both forward total memory and reverse total memory.
2. Set the total unit beforehand in the MEASURE SETUP mode (see Section 4.6.3.).
Total value
Preset value
Time
Reset
Reset
Reset
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1)
Typical operation Set the preset value to 100 m
3
.
* Set the total unit beforehand.
Keying order
Description
Indication
ź
Pressed twice to indicate "OUTPUT SETUP".
OUTPUT SETUP
ENT
ź
Pressed to indicate "ZERO ADJUST".
ZERO ADJUST
CLEAR
ź
Pressed 4 times to indicate "RANGE".
RANGE
FLOW RATE
ENT
ź
Pressed to blink the cursor.
RANGE
FLOW RATE
ź
Pressed to indicate "TOTAL" on 2nd line.
RANGE
TOTAL
ENT
ź
Pressed to indicate "TOTAL MODE".
TOTAL MODE
START
ź
Pressed 3 times to indicate "TOTAL PRESET".
TOTAL PRESET
0 m3
ENT
ź
Pressed to indicate the cursor.
TOTAL PRESET
0
00000000 m3
ź
Pressed 6 times to move the cursor.
* Note that input cannot be entered on the firsr digit (the
leftmost digit).
TOTAL PRESET
000000
0
00 m3
ź
Pressed to select "1".
TOTAL PRESET
000000
1
00 m3
Pressed to register it.
TOTAL PRESET
½½ COMPLETE ½½
Ļ
ENT
ź
ź
ź
ź
--- PRESET value registered.---
TOTAL PRESET
100 m3
ESC
Press
ESC
key 3 times, and
key 3 times to
resume the measurement mode.
0.000 m/s
0.000 m3/h
BULLETIN F-70