Page 69

background image

- 59 -

Typical operation Set the pulse value to 0.1 m

3

/pulse, and the pulse width to 50 ms.

* Set the total unit beforehand.

Keying order

Description

Indication

ź

Pressed twice to indicate "OUTPUT SETUP".

OUTPUT SETUP

ENT

ź

Pressed to indicate "ZERO ADJUST".

ZERO ADJUST
CLEAR

ź

Pressed 4 times to indicate "RANGE".

RANGE
FLOW RATE

ENT

ź

Pressed to blink the cursor.

RANGE

FLOW RATE

ź

Pressed to indicate "TOTAL" on 2nd line.

RANGE

TOTAL

ENT

ź

Pressed to indicate "TOTAL MODE".

TOTAL MODE
START

ź

Pressed to indicate "PULSE VALUE".

PULSE VALUE
1 m3

ENT

ź

Pressed to indicate the cursor.

PULSE VALUE

0

00000001 m3

ź

Pressed 7 times to move the cursor.

PULSE VALUE
0000000

0

1 m3

ź

Pressed 9 times to indicate the decimal point.

PULSE VALUE
0000000

.

1 m3

Pressed to register.

PULSE VALUE

½½ COMPLETE ½½

Ļ

ENT

ź
ź
ź
ź

--- “PULSE VALUE” registered.---

PULSE VALUE
0.1 m3

ź

Pressed to indicate the PULSE WIDTH.

PULSE WIDTH
5.0 msec

ENT

ź

Pressed to blink the cursor.

PULSE WIDTH

5.0 msec

ź

Pressed twice to select "50.0 msec".

PULSE WIDTH

50.0 msec

Pressed to register it.

PULSE WIDTH

½½ COMPLETE ½½

Ļ

ENT

ź
ź
ź
ź

--- “PULSE WIDTH” registered.---

PULSE WIDTH
50.0 msec

ESC

Press

ESC

key 3 times, and

key 3 times to

resume the measurement mode.

0.000 m/s
0.000 m3/h

BULLETIN F-70