Dwyer UXF2 User Manual
Page 69
- 59 -
Typical operation Set the pulse value to 0.1 m
3
/pulse, and the pulse width to 50 ms.
* Set the total unit beforehand.
Keying order
Description
Indication
ź
Pressed twice to indicate "OUTPUT SETUP".
OUTPUT SETUP
ENT
ź
Pressed to indicate "ZERO ADJUST".
ZERO ADJUST
CLEAR
ź
Pressed 4 times to indicate "RANGE".
RANGE
FLOW RATE
ENT
ź
Pressed to blink the cursor.
RANGE
FLOW RATE
ź
Pressed to indicate "TOTAL" on 2nd line.
RANGE
TOTAL
ENT
ź
Pressed to indicate "TOTAL MODE".
TOTAL MODE
START
ź
Pressed to indicate "PULSE VALUE".
PULSE VALUE
1 m3
ENT
ź
Pressed to indicate the cursor.
PULSE VALUE
0
00000001 m3
ź
Pressed 7 times to move the cursor.
PULSE VALUE
0000000
0
1 m3
ź
Pressed 9 times to indicate the decimal point.
PULSE VALUE
0000000
.
1 m3
Pressed to register.
PULSE VALUE
½½ COMPLETE ½½
Ļ
ENT
ź
ź
ź
ź
--- “PULSE VALUE” registered.---
PULSE VALUE
0.1 m3
ź
Pressed to indicate the PULSE WIDTH.
PULSE WIDTH
5.0 msec
ENT
ź
Pressed to blink the cursor.
PULSE WIDTH
5.0 msec
ź
Pressed twice to select "50.0 msec".
PULSE WIDTH
50.0 msec
Pressed to register it.
PULSE WIDTH
½½ COMPLETE ½½
Ļ
ENT
ź
ź
ź
ź
--- “PULSE WIDTH” registered.---
PULSE WIDTH
50.0 msec
ESC
Press
ESC
key 3 times, and
key 3 times to
resume the measurement mode.
0.000 m/s
0.000 m3/h
BULLETIN F-70