4.7.3. How to check the action of total pulses

Description
z

Checks the action of total pulse output.
The output action can be verified upon setting the number of pulses to be outputted per second.
Settable range: 1 to 100 pulses/s. (DO1 only)

Note: 1. The output pulse width is as currently set (see Section 4.5.6.1.)
Set the frequency taking the pulse width into account referring to the following expression.
Pulse count setting 1000/(pulse width [ms] u2)
Example: If the pulse width is set at 50 ms, select 10 pulses/s or less.
Note: 2. DO1 (transistor open collector) and DO2 (relay contact) operate simultaneously.

Before checking the action, confirm whether proceeding to an action is all right or not.

Note: 3. DO2 (relay contact) always operates at the rate of 1 pulse/sec.

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).

Typical operation Output 5 pulses/s.

Keying order

Description

Indication

ź

Pressed 4 times to indicate "MAINTENANCE MODE".

MAINTENANCE MODE

ENT

ź

Pressed to indicate "CURRENT CALIBRATION".

CURRENT
CALIBRATION

ź

Pressed twice to indicate "TOTAL PULSE".

TOTAL PULSE
1 PULSE/s

ENT

ź

Pressed to blink the cursor.
Note: Outputting simulated pulses starts.

TOTAL PULSE

0

01 PULSE/s

ź

Pressed twice to move the cursor.

TOTAL PULSE
00

1

PULSE/s

ź

Pressed 4 times to select "5".

TOTAL PULSE
00

5

PULSE/s

Pressed to register it.

TOTAL PULSE

½½ COMPLETE ½½

Ļ

ENT

ź
ź
ź
ź

--- 5 pulses/s registered.---

5 simulated pulses/s are outputted.

TOTAL PULSE
00

5

PULSE/s

ESC

ź

Pressed to stop the simulated output after the end of
checking the output.

TOTAL PULSE
005 PULSE/s

ESC

Press

ESC

key once, and

key once to resume the

measurement mode.

0.000 m/s
0.000 m3/h