How to check the action of total pulses – Dwyer UXF2 User Manual
Page 91
- 81 -
4.7.3. How to check the action of total pulses
Description
z
Checks the action of total pulse output.
The output action can be verified upon setting the number of pulses to be outputted per second.
Settable range: 1 to 100 pulses/s. (DO1 only)
Note: 1. The output pulse width is as currently set (see Section 4.5.6.1.)
Set the frequency taking the pulse width into account referring to the following expression.
Pulse count setting
1000/(pulse width [ms] u2)
Example: If the pulse width is set at 50 ms, select 10 pulses/s or less.
Note: 2. DO1 (transistor open collector) and DO2 (relay contact) operate simultaneously.
Before checking the action, confirm whether proceeding to an action is all right or not.
Note: 3. DO2 (relay contact) always operates at the rate of 1 pulse/sec.
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).
Typical operation Output 5 pulses/s.
Keying order
Description
Indication
ź
Pressed 4 times to indicate "MAINTENANCE MODE".
MAINTENANCE MODE
ENT
ź
Pressed to indicate "CURRENT CALIBRATION".
CURRENT
CALIBRATION
ź
Pressed twice to indicate "TOTAL PULSE".
TOTAL PULSE
1 PULSE/s
ENT
ź
Pressed to blink the cursor.
Note: Outputting simulated pulses starts.
TOTAL PULSE
0
01 PULSE/s
ź
Pressed twice to move the cursor.
TOTAL PULSE
00
1
PULSE/s
ź
Pressed 4 times to select "5".
TOTAL PULSE
00
5
PULSE/s
Pressed to register it.
TOTAL PULSE
½½ COMPLETE ½½
Ļ
ENT
ź
ź
ź
ź
--- 5 pulses/s registered.---
5 simulated pulses/s are outputted.
TOTAL PULSE
00
5
PULSE/s
ESC
ź
Pressed to stop the simulated output after the end of
checking the output.
TOTAL PULSE
005 PULSE/s
ESC
Press
ESC
key once, and
key once to resume the
measurement mode.
0.000 m/s
0.000 m3/h
BULLETIN F-70