- 37 -

(5) Setting the total pulse and preset value, and starting the total

Set the pulse value, pulse width and preset value.
Then, reset the total value to a preset value (factory set at 0), and start a total.

Proceed to the following from the measurement mode.

Keying LCD

indication/comment

key pressed twice.

1st line: [OUTPUT SETUP]

ENT

key pressed.

1st line: [ZERO ADJUST]

key pressed 4 times

1st line: [RANGE], 2nd line: [FLOW RATE]

ENT

key pressed.

2nd line blinks

key pressed.

2nd line: [TOTAL]

ENT

key pressed.

1st line: [TOTAL MODE], 2nd line [START]

key pressed.

1st line: [PULSE VALUE], 2nd line: [1m

3

] * As selected currently.

ENT

key pressed.

Cursor blinks on 2nd line.

and

key

pressed for composing a
pulse value.

Compose a desired pulse value. (See 4.5.6.1)

: Selects a numeric or decimal point.

: Shifts the place.

ENT

key pressed.

[**COMPLETE**] appears about 1 second on 2nd line, and then pulse value is
registered.

key pressed.

1st line: [PULSE WIDTH]. 2nd line: [5.0 ms]. * As selected currently.

ENT

key pressed.

Cursor blinks on 2nd line.

key pressed.

Select 5.0 ms, 10 ms, 50 ms, 100 ms or 200 ms. (See 4.5.6.1)

ENT

key pressed.

[**COMPLETE**] appears about 1 second on 2nd line, and then pulse width is
registered.

key pressed.

1st line: [TOTAL PRESET]. 2nd line: [0 m

3

]. * As selected currently.

ENT

key pressed.

Cursor blinks.

and

key

pressed.

Compose a desired preset value.

: Selects a numeric or decimal point.

: Shifts the place.

ENT

key pressed.

[**COMPLETE**] appears about 1 second on 2nd line, and then preset value is
registered.

key pressed 3 times.

1st line: [TOTAL MODE]. 2nd line: [START]. * As selected currently.

ENT

key pressed.

2nd line blinks.

key pressed.

2nd line: [RESET]. * Make sure beforehand total value can be reset.

ENT

key pressed.

[**COMPLETE**] appears about 1 second on 2nd line, and then total value is
reset.
2nd line: [STOP]. * Total stops.

ENT

key pressed.

Cursor blinks on 2nd line.

key pressed twice.

2nd line: [START].

ENT

key pressed.

[**COMPLETE**] appears about 1 second on 2nd line.
2nd line: [START]. * Total starts.

ESC

key pressed 3 times.

1st line: [OUTPUT SETUP].

key pressed 3 times.

Measurement mode is resumed.

BULLETIN F-70