beautypg.com

Wiring to each terminal, Bulletin f-70 – Dwyer UXF2 User Manual

Page 25

background image

- 15 -

3.3.4. Wiring to each terminal

Refer to the following diagram for carrying out wiring.

L

N

GND

GND

Iout(+) DO1(+) DO2(+) GND

HF1

NC

GND

RXD

TXD

+ −

GND

GND

SYNC

Synchronization RS-485

SHILD TXDR2 TXDR1

Iout(-) DO1(-) DO2(-)

GND

HF2

Power board terminal block

AC power source: 100 to 120 or 200 to 240 V AC,
50/60 Hz

Communication board terminal block (option)

External ground terminal

RS-232C

RS-485 and synchronization

DC power source: 20 to 30V DC

Power cable

Downstream
sensor cable

Upstream

sensor cable

Output signal cable

(analog output, DO1, DO2,

communication synchronization)

Ground terminal

Main board terminal block

Upstream sensor

Downstream sensor

Red (upstream sensor)

Black (downstream sensor)

Special signal cable (SX3C)

Iout : Analog output
DO1 : Transistor open collector
DO2 : Relay contact

Notes

1. All screws are M3 on the terminal block. Use crimp-style terminals for M3 and whose outer diameter is

‡5.8 or

smaller.

2. Be sure to connect to ground the power board terminal block or external ground terminal (class D ground).
3. For output signal, use multiple core cable as required.

BULLETIN F-70