How to validate the synchronization – Dwyer UXF2 User Manual

Page 97

background image

- 87 -

4.7.7. How to validate the synchronization

Description
z

Transmits ultrasonic waves according to synchronizing signals from the master flow transmitter.

z

Used to avoid the influence by mutual interference between several flow meters located near each other. Used, for
example, when they are installed near the same pipe, signal cables pass through the same conduit or there may
otherwise be signal interference or crosstalk.

z

Reception of synchronizing signal is checked (on slave flow transmitter).

Settable contents

MASTER,SLAVE

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).

Typical operation Select a SLAVE.

Keying order

Description

Indication

ź

Pressed 4 times to indicate "MAINTENANCE MODE".

MAINTENANCE MODE

ENT

ź

Pressed to indicate "CURRENT CALIBRATION".

CURRENT
CALIBRATION

ź

Pressed 6 times to indicate "SYNCHRONIZATION".

SYNCHRONIZATION
MASTER

ENT

ź

Pressed to blink the 2nd line.

SYNCHRONIZATION

MASTER

ź

Pressed to indicate "SLAVE".

SYNCHRONIZATION

SLAVE

Pressed to register it.

SYNCHRONIZATION

½½ COMPLETE ½½

Ļ

ENT

ź
ź
ź
ź
ź
ź
ź

--- SLAVE registered.---

* The reception of synchronizing signal is checked by

the SLAVE, if registered. "NORMAL" appears if the
synchronizing signal has been received, or "ERROR"
if not received properly.

Note: Nothing appears if "MASTER" was registered.

SLAVE CHECK
NORMAL

ESC

ź

Pressed to indicate "SLAVE".

SYNCHRONIZATION
SLAVE

ESC

Using

ESC

key and

key, resume the

measurement mode.

0.000 m/s

0.000 m3/h

BULLETIN F-70